In the realm of sophisticated maritime equipment, the Gulfstream 400 Spectra stands out as a premier watermaker, designed to enhance life on board by providing a reliable source of fresh water. Crafted in Sausalito, California, this system embodies innovation through its detailed operation procedures, including priming, purging, system startup, and maintenance routines. The unit's meticulous design requires the opening of a pressure relief valve and the individual running of its pumps to prime, ensuring that each step, from system startup to storage, is executed with precision for optimal performance. Emphasizing the importance of proper installation and maintenance, the manual offers a comprehensive guide encompassing everything from basic operation, such as attaching the product service tube and setting valves, to regular maintenance protocols like backflush cycles and pre-filter servicing. Detailed instructions on storage recommend attaching service hoses to a clean container and initiating a backflush cycle, followed by the incorporation of a specific storage chemical, highlighting the system's need for regular care to prevent damage. Moreover, the manual serves as a testament to Spectra Watermakers' commitment to customer satisfaction and product quality, ensuring long-term reliability and efficiency of the Gulfstream 400 system. This dedication is further illustrated by the inclusion of service valves in the installation kit, aimed at simplifying the pickling and cleaning processes, thereby enhancing the overall user experience.
Question | Answer |
---|---|
Form Name | Gulfstream 400 Spectra Form |
Form Length | 38 pages |
Fillable? | No |
Fillable fields | 0 |
Avg. time to fill out | 9 min 30 sec |
Other names | gph, gulfstream 400 spectra watermakers, 2000, SC-2 |
Feed Pressure |
|
Pressure Relief Valve |
Product Flow |
Open 1/2 to 1 turn only |
|
SPECTRA |
GULFSTREAM 400 |
|
|
Sausalito, California Phone (415) |
Fax (415) |
Pump 1 |
Salinity |
Backflush |
|
||
Pump 2 |
Reject |
Good |
|
Basic Operation:
Priming
Open pressure relief valve, run each pump individulally to prime system. Pumps will not prime simultaneously.
Purging
If system has been stored, open pressure relief valve and run for 20 minutes. Attach the product service
tube and place into a container. Set product service valve to "service" position, close pressure relief valve and run for 20 minutes. Discard product water.
Startup
Set all valves to "run" position and close pressure relief valve. Start one pump for efficiency, two pumps
for higher production. Check for the "Good" salinity indicator to light. Monitor feed pressure for
NormalOperation
For best performance; run system to fill tanks then turn off pumps. Have the domestic water system perssurized then start backflush cycle.
Storage
Attach service hoses to unit and lead to a clean container. Open pressure relief valve and start backflush cycle. After backflush cycle, set brine discharge valve to "service" position and then
Allow 1 gallon of flush water into the container and then set the discharge valve to "run". After backflush cycle, mix one container of
and run for 10 minutes, recirculating the solution. Set the brine discharge valve to "run" to empty container. Turn off pumps, return all service valves to "run" position. Service strainer and
Purge system on next start up.
Intake Service Valve |
Product Service Valve |
Brine Discharge Service Valve |
Run |
|
Run |
Off |
Run |
Off |
|
Service |
|
Intake Service |
|
Brine Discharge Service |
Hose Connection |
|
Hose Connection |
|
Product Service Tube Connection |
|
GULFSTREAM 400
OPERATION AND INSTALLATION
MANUAL
Spectra Watermakers
A Division of Edinger Marine Service, Inc.
REVISED 10/15/99
20 Mariposa Road |
San Rafael CA 94901 |
Phone 415.526.2780 |
Fax 415.526.2787 |
page 1
Introduction
Thank you for choosing a Spectra Watermaker! It is our wish for you to be completely satisfied with our product. We have inspected and tested all of the components before shipping, and feel we have produced one of the finest products available.
Your long term satisfaction with this system will depend greatly on how the Spectra system is installed. The unique performance is obtained not only by our new pressurization process, but by careful attention to reducing flow restriction, mechanical friction and electrical resistance. All of this is affected by the final installation, and that is ultimately up to you. By reading, understanding and following this manual, even if you are not doing the installation yourself, we are confident you will be happy with your new Spectra system.
With years of experience installing and repairing many different kinds of watermakers, we have seen nearly everything that should or should not be done. This manual includes the summation of that knowledge, tailored for the Spectra Watermakers.
It is very important that you follow the instructions in this manual. They are designed to help you "customize" the installation to your boat, while maintaining performance. If any questions arise, please call us for help immediately. This is a unique product so be careful of "experts," as they most likely will not have gone through the same learning curve as we have during the development of the Spectra systems.
If someone else is installing the watermaker for you, review all aspects of the system with them before the drilling and cutting begins. This will help avoid making mistakes. If you have any doubts, please call us. We will be happy to help tailor the system to your particular needs. As with any sophisticated device, there may be “bugs” or minor problems. Please call us for consultation or advice before attempting to repair a new system.
"Use it or lose it" really does apply to watermakers, but we prefer "use it or pickle it". All watermakers will suffer damage if they are run and left without storage treatment or "pickling". We have found that if the pickling and cleaning processes are made easy there is a higher probability of it happening. This is why we have included service valves in the Installation Kit. Even though the system can work without them, they are very helpful in the servicing of your system.
In conclusion, having an adequate and reliable watermaker can improve the quality of life on board more than almost any other piece of equipment. It is our sincere desire that you will be able to take full advantage of your new Spectra Watermaker, confirming that you have indeed made the best choice! And don’t forget to tell all your friends!
Thank you from all of us at Spectra Watermakers.
page 2
Table of Contents
System and |
page 4 |
Shipping and Shopping Lists |
page 5 |
Features |
page 7 |
Specifications |
page 8 |
Component Description |
page 9 |
Operating Procedures |
page 13 |
Shutdown Procedures |
page 16 |
Maintenance |
page 18 |
Membrane Cleaning Procedures |
page 19 |
Membrane replacement |
page 21 |
Installation Basics |
page 22 |
General Plumbing |
page 23 |
Component Placement |
page 25 |
Clark Pump Diagram |
page 27 |
Component Dimensions |
page 28 |
Plumbing Diagram |
page 29 |
Plumbing Detail Diagram |
page 30 |
Clark Pump Mounting Diagram |
page 31 |
High Pressure Tube Instructions and Diagram |
page 32 |
1/4” Tube Fitting Instructions Diagram |
page 33 |
Mounting Templates |
page 34 |
Electrical |
page 35 |
Wiring Diagram |
page 36 |
Troubleshooting |
page 37 |
Warranty |
page 38 |
page 3
System and
System Information
Model Number |
________________ |
Serial Number |
________________ |
Date Purchased |
________________ |
Purchased From |
____________________________ |
|
____________________________ |
|
____________________________ |
Installation Date |
________________ |
Measure after 3 and 24 hours of pressurized run time in similar conditions.
|
3 hrs |
24 hrs |
Feed Water Temperature |
__________ |
__________ |
Battery Voltage |
__________ |
__________ |
Amperage Draw |
__________ |
__________ |
Feed Pressure (psi) |
__________ |
__________ |
Product Water Flow (gph) |
__________ |
__________ |
page 4
Shipping List
Standard System
(1)Operation and Installation Manual
(1)Clark Pump
(1)Membrane and Housing Assembly
(1)Main Pump Module including: Pumps, heat sinks, and cooling fans Sea water strainer Accumulator tank
Monitor and control panel Automatic salinity control system Automatic fresh water backflush system 20 and 5 micron
(1)Remote Control Panel and Wiring Harness
(1)Installation Kit Including:
Clark Pump mounting hardware Membrane housing assembly mounts (10’) high pressure tube
50’ of 5/8” hose
25’ of 1/4” product tubing Hose and tube fittings Hose clamps
One container of
Options
Clark Pump Mounting Bracket
Basic Cruise Kit
Includes three
chemical containers
Offshore Kit
Includes Clark pump seals, tools, two
Spare Feed Pump
Hand held salinity monitor
page 5
Shopping List
In addition to the installation kit.
Refer to manual for wire size and type of fasteners
Fasteners/Hardware for mounting components to desired surfaces. Marine grade duplex wire, connectors and wire ties with mounts. One 2 gallon plastic mixing container.
One 25 amp fuse or circuit breaker switch
Tools Required
Wrench set to 7/8”
Adjustable wrench
3/8” drill, drill index and a hole saw set Jig saw or coping saw Screwdrivers
5/16” nut driver for hose clamps Wire cutter
Wire stripper
Electrical crimping tool for insulated connectors
page 6
Features
The Spectra Gulfstream a full featured,
Using the patented Clark Pump and paying careful attention to system efficiency, the Spectra Gulfstream, in the efficiency mode, operates on a fresh water to power ratio that surpasses everything in the marine industry. The system makes an average of 9 gallons per hour on 9 amps @ 12.5v DC, less than 13 watt hours per gallon! That means that making one gallon of fresh water is like running only a small DC cabin light for one hour! In high production mode, the Spectra Gulfstream produces up to 17 gallons per hour using less than 20 amps.
The
The production of fresh water is stable through the system’s normal operating sea water temperature and salinity range. Only the working pressures change with the water conditions. A typical leading brand 400 gallon per day unit running in 61 degree F water will only make 280 gallons per day. The Spectra Gulfstream will still be producing up to 400 gallons per day on a fraction of the power.
The automatic salinity control system, automatic fresh water flush system, and a full function remote control panel are included standard
Pressures are
Self priming
A
The system’s noise level is exceptionally low and unobtrusive, especially compared to other systems of similar capacity. The feed pumps have a smooth, steady sound and almost no vibration.
An installation kit is included.
page 7
Spectra Gulfstream Performance Specifications:
Efficiency Mode (one feed pump)
Sea water temperature @ 50 degrees F (10 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
1.6 gpm (6.0 lpm) |
82 psi (5.6 bar) |
685 psi (46.6 bar) |
9.6 |
9.3 gph (36.0 lph) |
Sea water temperature @ 77 degrees F (25 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
1.6 gpm (6.0 lpm) |
68 psi (4.6 bar) |
570 psi (38.8 bar) |
8.5 |
9.5 gph (35.2 lph) |
Sea water temperature @ 90 degrees F (32 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
1.6 gpm (6.0 lpm) |
68 psi (4.6 bar) |
550 psi (37.4 bar) |
8.4 |
9.5 gph (35.9 lph) |
High Production Mode (two feed pumps)
Sea water temperature @ 50 degrees F (10 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
2.3 gpm (8.7 lpm) |
110 psi (7.5 bar) |
880 psi (59.8 bar) |
21.2 |
15.0 gph (56.8 lph) |
Sea water temperature @ 77 degrees F (25 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
2.6 gpm (9.8 lpm) |
98 psi (6.6 bar) |
700 psi (47.6 bar) |
19.6 |
16.5 gph (62.4 lph) |
Sea water temperature @ 90 degrees F (32 degrees C):
Voltage |
Feed flow |
Feed pressure |
High pressure |
Amperage |
Product |
13.8v |
2.6 gpm (9.8 lpm) |
98 psi (6.6 bar) |
665 psi (45.2 bar) |
19.1 |
17.0 gph (64.3 lph) |
Sea water salinity @ 35,000 parts per million TDS
Salt rejection: 99.0% ~99.4% Typical purity test: < 400 mhos, 200 ppm, 99.4% pure Performance Tolerance +/- 10%
Maximum Operating Temperature 113 degrees F (45 degrees C)
Dry weight: 80.5 lbs.
Component Description
page 8
Clark Pump
This is the heart of the system. The Clark Pump uses two opposing cylinders and pistons that share a single rod through a center block. A reversing valve, controlled by a pilot valve that is actuated by the pistons, allows the cylinders to alternate between driving and pressurizing. Pressurization is achieved by the rod being driven into the other cylinder. As the rod enters the cylinder, it pushes the piston which circulates the water out through the membrane and back to the reversing valve. The valve directs the water back into the same cylinder behind the piston, creating a closed loop. The rod displaces water as it enters the cylinder, raising the pressure enough for reverse osmosis to occur in the membrane, thus fresh water equal to the volume of the rod is produced on every stroke. The driving cylinder has feed pressure pushing on the top of the piston which has ten times the surface area than in the end of the rod. The water under the piston (brine) is discharged overboard. This gives the piston a
The speed of this process is determined by the water flow from the feed pump. The feed pump has a nearly constant flow rate at a given voltage so the production of fresh water is nearly constant, only the pressures change with changes in sea water temperature and salinity. (see "Membrane" section)
A pressure relief valve and a reset button are incorporated in the Clark Pump to depressurize the system for priming and service and resetting the reversing valve if the Clark Pump is inadvertently stopped with the pilot and reversing valves at dead center.
Reverse Osmosis Membrane and Housing
The “RO” membrane element does the real work. A
In the efficiency mode, approximately 85 gph of high pressure salt water flows across the membrane layer as 8.5 gph of fresh water is forced through the membrane and lead by a capillary layer to the catch tube. The excess salt water flowing out of the membrane is approximately 1/10th saltier and is called “brine.” No salts are collected in the membrane. Membrane function is affected by sea water temperature and salinity. The warmer or less salty the water, the easier it is for the fresh water to pass through the membrane. This means that since the water production of the Spectra systems are nearly constant, warmer water lowers the operating pressures. The membrane is carefully sized to handle typical sea water temperatures and salinity, keeping the pressures within limits.
Component Description, cont.
Main Pump Module
*The following 9 items are
page 9
Note: The main module components are
*Dual Feed Pump System
The system is simply two identical self priming,
*20 and 5 Micron Dual
The
*Accumulator Tank
The accumulator tank is an air chamber divided by a rubber diaphragm. The top chamber is pressurized with air to 10 psi below the expected feed pressure. For example, if the feed pressure is 70 psi, the tank should be pressurized to 60 psi when the system is off. When running, feed water will enter the lower chamber until the pressures equalize. Its function resembles a cushion to smooth out the feed pressure spikes when the Clark Pump “shifts.” This will extend the life of the feed pump and quiet the shifts between strokes.
*Feed Pressure Gauge
The feed pressure gauge is an indicator of system operational status. Higher than normal pressures indicates the need for
Component Description, cont.
*Product Water Flow Meter
The product water flow meter indicates the flow in gallons per hour. The reading is taken at the middle of the indicator ball.
page 10
*Power Control Module
The Spectra power control module is a plastic enclosure that is mounted inside on the lid of the main enclosure. The power feed wires connect to the face. It has all of the power relays, fuses, and electronics for the feed pumps, the salinity control system, and the automatic backflush system. The module is connected to the ships power from a single breaker that is rated for all the system loads. The feed pumps, salinity and backflush solenoid valves, salinity probe, and the remote paned are all wired to the control module.
*Service Valve Manifold
The service valves are in the feed, brine discharge, and product lines to provide easy access for proper maintenance. The service hoses attach to the panel face and are lead to a cleaning container. IMPORTANT: The intake and discharge service valves will close off all the ports if the lever is moved to the “Off” position. The product service valve cannot close the ports when the lever is moved half way. This precludes the possibility of blocking the product flow and causing an over pressure condition that could burst the tube. The salinity control solenoid, the backflush solenoid, and the feed water check valve are incorporated in the manifold.
*Salinity Control System
The salinity control is integrated with the power control module and controls a product water diversion valve mounted in the service valve manifold. There are “reject” and “good”
salinity indicator lights on the main and remote control panels. The salinity control continuously samples the purity of the product water. If the probe detects
*Automatic Fresh Water Backflush Systems
Warning: The ship’s domestic water system must be pressurized before flushing
An automatic fresh water backflush system allows the operator to start a timed backflush cycle from the control panels. The automatic system incorporates a charcoal filter to remove chlorine, an electric solenoid valve, a check valve mounted in the service manifold, and a power relay to run one feed pump. When the timer is initialized the valve opens to let pressurized charcoal filtered fresh water into the feed system and one feed pump starts. The check valve prevents the fresh water from going out the intake
The pressure relief valve does not have to be opened during purging because the Spectra’s unique high pressure system will not generate membrane damaging excessive product flow when brackish or fresh water is introduced into a pressurized system.
Component Description, cont.
* Remote Pressure Relief Valve
The pressure relief valve on the panel face duplicates the function of the pressure relief valve on the Clark Pump. It is plumbed to a SS fitting on the top of the Clark Pump with high pressure 1/4” tubing. The high pressure 1/4” tubing is different than the product 1/4” tubing
page 11
and should not be interchanged. Opening the valve 1/2 turn will allow the system to be run without developing pressure. This is necessary for priming and membrane cleaning.
Remote Control Panel
The Spectra remote control panel is integrated with the Spectra power control module, salinity control system, and the automatic back flush system. The panel has switches for both feed pumps, a switch to start a backflush cycle, and lights that indicate when a feed pump is on, the backflush cycle in in progress, product water is being rejected, and when good product water and going to the tanks. The remote panel is wired to the power control module by a waterproof connector on a
Sea Water Strainer
The strainer filters the salt water to the feed pump. It has a cleanable, fine mesh, stainless steel element. The filter mesh is sized to stop debris that could jam in the pump’s check valves. A coarse strainer should not be substituted.
Installation Kit
The installation kit includes 50 feet of 5/8” reinforced
Mounting Bracket (optional)
The mounting bracket is a
Basic Cruise Kit (optional)
The kit includes three
Offshore Kit (optional)
The kit includes all seals and
Operating Procedures
New System
Caution: Avoid running the system if the vessel is in contaminated water, such as in a harbor or canal. Membrane damage and
page 12
the pressure relief valve open and again 5 times with the pressure relief valve closed with the product water diverted. This will flush the storage solution from the membrane. The system can now be run using a clean source of sea water in the cleaning container. Unchlorinated fresh water mixed with sea salt can also be used. Use the intake and brine discharge service hoses to recirculate the salt water from the container.
1.Check battery voltage. A nominal 12.8v DC to 14.4v DC should be indicated.
2.Confirm that the intake
3.Have the intake and brine discharge service valves in the normal run position and the product service valve in the “Service” position. Direct the product service tube to a container or into the bilge. The pressure relief valve on the Clark Pump should be closed.
4.Open the pressure relief valve on the panel face 1/2 turn and start one feed pump from the main or remote control panel. Listen for the feed pump to prime itself and the Clark Pump start to “shift.” Check the brine discharge for water flow. Shut down the first pump and start the second pump. After the second pump is primed, the first pump can be restarted. Check the feed pressure, about
5.Check the feed pump high pressure cut out switch set point. Do one pump at a time. With the system primed and running unpressurized, block the brine flow by moving the brine discharge service valve lever to the “Off” position. This will stop the water flow and cause the feed pump to shut off. Check the feed pressure gauge, it should be holding steady at about 115 to 125 psi. The pressure can be adjusted by tightening (higher pressure) or loosening (lower pressure) the Allen head set screw on the pump switch using a 5/64” Allen wrench. Repeat for the second pump. Now is a good time to check for leaks in the feed and brine connections as they are now pressurized to maximum.
6.Run the system unpressurized for 30 minutes after
7.Close the pressure relief valve to pressurize the system and watch for the feed pressure to rise. If the pressure does not rise, check that the pressure relief valve on the Clark Pump is closed.
than 5 to 10 lbs. over normal when the Clark Pump shifts and should stay under 85 psi at all
Operating Procedures, cont.
times in the efficiency mode and under 100 psi in high production mode. The feed pumps should never be allowed to cycle their high pressure cut out switches on and off during the shift. The pump sound should remain almost steady. If the pressure spikes are high and/or the pump stops momentarily during the shift, check the high pressure cut out set point and/or check the accumulator
page 13
off even when all else is normal. In those conditions only run one feed pump.
8.After one hour, the product service valve can now be switched the “Run” position to direct the flow to the water tank. The “Good” salinity indicators on the panels should light. Break open the product line to the tank to see if the water is flowing to confirm proper operation of the diversion valve.
9.At three hours of pressurized run time, and again at 24 hours,
Normal
1.For normal
2.If air is in the system and one or both feed pumps have lost their prime, open the pressure relief valve and run one pump at a time to purge the system. The feed pumps cannot prime if their output is under any pressure. Remember to purge the pumps separately.
3.After starting and pressurizing the unit, check for leaks and wait for it to stabilize, then check for proper readings (see Specifications, page 8 ). Air in the system will cause erratic shifts and readings until it is purged out of the system. This may take up to 10 minutes.
Warning!
On start up, if the feed pressure keeps rising until the feed pumps shut off and the Clark Pump is not shifting, depressurize the system and press the reset button on the side of the reversing valve on the Clark Pump, all the way in, to unstall the system. To avoid a stall condition, always turn off the system right after the Clark Pump “shifts”. Also check that the service valves are in the normal run position.
Pickled System
If the system has biocide in it (also known as “pickling” or storage solution), open the pressure relief valve 1/2 turn and run the system unpressurized for 30 minutes to purge the biocide. Attach the product service line lead to a container or into the bilge. Switch the product service valve to the “Service” position to manually reject any product water. Close the pressure relief valve to pressurize the system. Run for another 30 minutes then switch the product service valve back to the “Run” position.
Operating Procedures, cont.
Unknown Condition
If the system has been left unpickled for an extended length of time or the history is unknown, perform this simple test before running the system:
Open the
page 14
brine that first comes out. If the water is discolored and smells bad, perform an
Normal Operation
Spectra Watermakers are designed to run continuously with only
In moderate climates, if the membrane is mounted in a cool place, it is not necessary to flush the system every time if the unit is used once every one to two days. Flush the system if it’s not going to be operated for more than two days and up to one week. Flush the system after every use in hot conditions. Pickle the system if longer periods of
Running the unit during charging cycles will produce more water because of the higher voltage (see Performance Specifications). System readings should be periodically checked and compared to the
The system’s automatic salinity control manually diverts any brackish product water into the brine discharge and the “Reject” light will illuminate. When potable water is detected, the “Good” light will illuminate and the product will be directed to the tank.
Use the high production mode (both feed pumps running) when the batteries are under
charge.
Use the efficiency mode (one feed pump running) when the batteries are not being charged.
If too much air is introduced into the system when the system is running and pressurized, the feed pumps will cavitate and water flow will stop or only one will cavitate and the production will be fall to the efficiency mode level. This will not damage the system. Open the pressure relief valve and reprime the feed pumps separately.
All reverse osmosis watermakers are affected by water temperature and salinity. The Spectra operating pressures will rise with colder temperatures and/or higher salinity, and fall with warmer temperatures and/or lower salinity, but the product flow is affected very little.
page 15
Shutdown Procedures
When not in use, biological growth in the membrane is the leading cause of membrane fouling. A warm environment will cause faster growth than a cold environment. The following procedures assume a warm environment, such as in the Tropics.
Normal Shutdown,
Run system to fill the tanks. Check system for normal operation. If OK, listen for the Clark Pump to “shift” then immediately turn off the power. There is a small chance that the Clark Pump will be left in a stalled condition if it is shut down during a “shift”. Make sure that the ship’s pressure water system is on, then start the flush cycle with the backflush momentary switch on the main or remote panel. When the system turns itself off, service the
On the next start up, if the Clark Pump is found to be stalled, turn off the system, open the pressure relief valve, and push the reset button on the right side of the reversing valve in completely, then restart. It is good practice to close the intake
Storage Shutdown or "Pickling,"
The Spectra
The
1.Attach the service hoses to the face of the unit and direct them to at least a clean 2 gallon plastic or SS mixing container. With the pressure relief valve open, start the backflush cycle. After the backflush ends, turn the brine discharge service valve to the “Service” position. Restart the backflush cycle, and allow one (1) gallon of product water to flow into the container then turn the brine valve back to “Run”. Let the backflush stop automatically or reset it by turning off the main power. Do not use tap water or chlorinated water!
If the automatic backflush system is not functional, perform a manual backflush for pickling in the following manner. Attached the service hoses and lead them to a 5 gallon container. Close the pressure relief valve and start the system normally. When the green “good” indicator light comes on, turn the product service valve to the service position. Fill the 5 gallon container with product water then shut the system down. Switch the intake service valve to the “service” position. Open the pressure relief valve and start the system. Allow four gallons of product water to flush the system, keeping one gallon in the container. Continue with step # 2.
Shutdown procedures cont.
page 16
2. When the back flush cycle stops and one gallon of fresh water is in the container, mix one container of the
Do not let skin, eyes, or lungs have any contact with the storage and cleaning chemical.
If the solution becomes excessively discolored, mineral scale and live biological growth may be present and membrane cleaning is recommended. Refer to the membrane cleaning section for further instructions.
3.After 10 minutes, turn the brine discharge valve to the “Run” position. When the container is empty, shut off the system.
4.Close the intake
5.Remove and clean the sea water strainer screen and
6.When returning the system to service after pickling, check that the intake
7. To winterize, Open the pressure relief valve and flush 5 gallons of potable water system antifreeze through the system instead of using the normal
page 17
Maintenance
The Seawater Strainer
The sea water strainer’s stainless steel element should be inspected, removed and cleaned as needed. Be careful to ensure that the
The Micron
Service the
Used pleated elements can be carefully cleaned if they are not too compacted. Do not use full pressure from a garden hose or scrub with a hard brush on the pleats. Gently spray the outside with water and then plug one end with your hand, filling from the other end to flush from the inside. Be careful to not distort the pleats. Another way is to tow them behind the boat until clean, 4 knots maximum. Slip a line through the filter and tie a stopper knot. Elements that show any signs of deterioration should be replaced. Check for leaks on the first run after a filter service.
The Charcoal Backflush Filter
Replace the charcoal filter element every 6 months.
The Membrane
The membrane needs to be cleaned only when it has lost up to 10% of its capacity due to fouling or the product quality degrades. The leading cause of fouling is from biological growth that occurs when the system is left unused for long periods without flushing or pickling. Fouling from mineral scaling can happen under certain sea water conditions, and from rust. Poor tasting water or the “Reject” light on continuously and/or higher than normal pressures and amp draw for the sea water temperature and salinity conditions are the main indicators of membrane fouling. Other conditions can cause high pressure (such as in very cold water). Low product flow is usually due to low voltage or a clogged
There are two types of cleaners; acid and alkaline. The acid cleaner
If the membrane fails to respond to both cleanings, this is an indication of another problem with the system, or that it is time to replace the membrane. Call Spectra Watermakers before removing a membrane.
page 18
Membrane Cleaning Procedures
For normal cleaning, the
If the history of the system is unknown or has been left “unpickled” for an extended length of time and biological growth is present, it is recommended that the system is cleaned with the
Cleaning Procedures
Note: procedures are the same for the
The
1.Attach the service hoses to the face of the unit and direct them to heated container. Make sure the domestic water pressure system is on then start the automatic backflush system. When the flush cycle stops, move the brine discharge to the “Service” position and restart the flush cycle. Allow one gallon of flush water to flow into the container, then move the brine discharge valve back to “Run”. Do not use tap water or chlorinated water! Distilled or charcoal filtered water is acceptable if using the flush water is not an option.
3.Open the bypass valve one half turn, this is important! Mix 1 Spectra chemical container into the 1 gallon in the container Do not let skin, eyes, or lungs have any contact with the storage and cleaning chemical.
3.Place the intake, brine discharge, and product service valves in the “Service” position. Start heating the water.
4.
again Membrane cleaning, cont.
for 1 hour while heating the solution, then turn off for one hour. Do not let the water get above 135º F!
page 19
5.Turn the brine discharge service valve to the “Run” position, then run the system until the container is empty.
6.Place the intake service valve to the “Run” position. Run the system unpressurized for at least 30 minutes to purge the cleaning solution.
7.After purging, close the pressure relief valve to pressurize the system. Manually divert the product water with the product service valve into the cleaning container. After 30 minutes, check the feed pressure, product flow, and quality, to determine if the membrane is functioning normally. If the cleaning solution became very discolored during the cleaning process, it is recommended to repeat the process. If the alkaline cleaning did not affect the membrane enough, repeat the cleaning process using the acid cleaner. Always manually reject the product water for 30 minutes after cleaning.
8.If the membrane performance is still poor and no other problem can be found, contact Spectra Watermakers for further assistance.
page 20
Membrane Element Replacement
Do not remove the membrane from the housing unless absolutely necessary. Troubleshoot for all other causes of poor performance and try the recommend procedures first.
Caution: membranes must be kept wet to prevent permanent damage.
Disassembly:
Disconnect the high pressure tubes, cut the housing mount bands, and move the membrane housing to a work area that can handle water spillage. Clean off any deposits on the ends with a stiff brush and fresh water. Note which end goes where so they can be reinstalled the same way later.
Loosen the retainer rings with an adjustable deck fill fitting wrench or use a hard plastic or wood punch and a mallet on the slot provided. As the retainer rings unscrew, twist and pull the end fittings to follow the rings out. This way they can be removed with the factory plumbing fittings in place. By first removing the plumbing fittings, the rings can be removed and then the end plugs removed from the housing. Inspect the end plugs and
Push the membrane element out of the housing by first pushing on the outlet end. If it resists, try pushing on the other end. If the membrane is going to be kept for any reason, seal it in plastic before it dries out. Do not open the new membrane packaging until ready to install.
Assembly:
Install the inlet end plug first. Lubricate the
Install the outlet end plug the same way as the inlet. It will go in harder because of the product tube
A new membrane has a pickling solution in it and should be thoroughly flushed before use. Follow the new system start up procedures.
Open the pressure relief valve 1/2 turn and start the system. Look for leaks at the end plugs. If a large leak is present, the large
page 21
Installation Basics
The Spectra Watermakers are designed for energy efficiency. All aspects of the system have been optimized for this goal, therefore any changes in the basic layout or components will most likely reduce this efficiency. The overall length of the feed, discharge and high pressure hoses have little effect on performance and can even help smooth out the Clark Pump shift pulses but that every hose bend or 90 degree fitting absorbs power. Avoid adding 90 degree fittings. Use gentle hose bends where possible. A good installation will be a proper compromise between minimum hose and wire runs, accessibility, environment and esthetics.
It is recommended that a designated 3/4” intake
Do not install the intake close to or down stream of a head discharge. Install as far below the water line and as close to center line as possible to avoid contamination and air induction.
We recommend that a reputable boatyard install any underwater
The discharge
Double clamp all hose connections below the water line if possible.
Fittings
Pipe fittings that are to thread into plastic should be coated with at least 3 wraps of Teflon tape. To avoid cracking the plastic, tighten until only 1/2 of the threads are buried. This may not seem tight, but they will not, and should not, bottom out like a metal to metal connection.
Metal to metal pipe connections should have
Be very careful to avoid getting any tape over the end of the fitting that might get into
page 22
the system. To insure this does not happen, leave the very first thread uncoated.
General Plumbing
Follow the Clark Pump and Plumbing Diagrams on pages 27, 29, and 30 for hose and tube connections. All feed and discharge hoses are made of 5/8" reinforced nylon. The high pressure tubing is 1/2” OD connected with 1/2” tube to pipe 316 SS compression fittings. The 90 degree high pressure SS tube fittings mounted in the high pressure in and out on the Clark Pump can be rotated up to 1/2 turn for better tube runs. Straight SS tube fittings can also be used. The product water lines are 1/4" black plastic tubing with connectors. Feed water operating pressure up to 125 psi requires vigorous hose clamp tightening. It is recommended to lubricate the clamp screw and use a nut driver for the final tightening. Double clamp all connections that are below the water line on the suction line to the feed pump.
If you use feed hose or product tubing that is not supplied by Spectra Watermakers, it is mandatory to use products with a minimum burst pressure of 250 psi. If other types of pre- filter housings are used, they need to rated at 125 psi minimum working pressure.
Feed and Discharge Hoses
Route the 5/8" hose smoothly from the intake
High Pressure Tubes
Be sure to follow the Clark Pump and Plumbing Diagrams for proper connection between the Clark Pump and the membrane housing. Membrane water flow direction is marked on the housing. Rotate the 90 degree high pressure tube fittings on the Clark Pump for ideal tube runs. The high pressure tubes are usually made in place. Loosely route the tubes between the Clark pump and the membrane and cut long. Minimum tube bend radius is 6”. Follow the tube connection instructions on page 32. Connect the tubes to one of the components, secure the tube runs, and then trim and connect to the other component. A 90 degree bend in a hose is better than a 90 degree fitting. A tube, when mounted, should have at least one gentle bend to allow for expansion. Do not connect a tube straight between hard mounted fittings.
If other than the Spectra supplied tube and fittings are used, they need to be at least 1/2" ID if hose, 1/2” OD if tube, with
When connecting the tubes to the components, be sure to hold fitting body during the final tightening.
General Plumbing, cont.
page 23
Remote Pressure Relief Valve Tubing
A coil of high pressure 1/4” plastic tubing is
The product 1/4” tubing is not rated for high pressure and cannot be used for the pressure relief valve system.
Product Water Tubing
The product water lines are 1/4" plastic tubing. Minimum tube bend radius is 3.” Connection can be made to either end of the membrane housing. Refer to the tube fitting assembly diagram on page 33. IMPORTANT: The product water flow should not be restricted in any way. If the line gets blocked, the pressure will build until the line fails. Restricting the flow will reduce efficiency and may cause damage to the membrane. Connect to the water tank at or near the top. The ideal spots are to tee into the tank vent line or the fill hose. Do not tap into the bottom of the tank or a water feed manifold. Air can be pulled through the membrane and affect the domestic water system.
Fittings
Pipe fittings that are to thread into plastic should be coated with at least 3 wraps of Teflon tape. To avoid cracking the plastic, tighten until only 1/2 of the threads are buried. This may not seem tight, but they will not, and should not, bottom out like a metal to metal connection.
Metal to metal pipe connections should have
Component Placement
page 24
Refer to the Dimension & Plumbing Diagrams on pages 28 & 29
Main Pump and Control Module
The main pump module is designed to mount on a horizontal surface that can be up to 4’ above the waterline. Locate in an area that allows easy access to the face panel and the service hoses can reach a container placed in a spot where water spillage is acceptable. Allow for a minimum of 2” of clearance on the sides and 13” over the top, and at least 1” of clearance at the back. Extra space over the top is preferred for easier internal access through the hinged top. Use the unit itself as a template for drilling the mounting bolt holes. Use quarter inch bolts or lag screws with washers. If lag bolts are used they must thread into at least 1” of solid material. Drill pilot holes for lags bolts.
Clark Pump
Refer to the the Clark Pump Mounting Diagram on page 31 to install the mounting brackets and shock mounts. Bolt the mounting brackets to the base of the center block with the short 5/16” SS bolts and washers, sandwiching the plastic washers in between the center block and the brackets. The brackets will over hang from the front and back of the center block. Moisten the rubber shock mounts and slip them into the bracket holes then install the plastic inserts.
The Clark Pump can be mounted on any horizontal surface that will support it’s weight and is not subjected to excessive vibration. Height above or below the water line is unrestricted. Mounting on a vertical surface can be expedited by using the optional "L" mount bracket. For hanging from overhead, use two of the "L" brackets. Avoid mounting in the middle of a large unsupported flat surface that could amplify the sound of the pump shift. The Clark Pump can be mounted in any position if it is hard mounted without using the brackets. More noise may be transferred to the vessel.
Allow room in the front and back for access to the high pressure tubes and hose connections. If the space behind the unit is restricted, attach the high pressure tubes to the Clark Pump before final mounting. The 90 degree fittings supplied are acceptable for the feed and high pressure tube connections to the unit and can be rotated for easy tube runs. Straight fittings can also be used. The brine discharge hose connection to the reversing valve body should be straight if possible, and can be connected to either end of the reversing valve. Plug the unused port.
Make sure that the area around and under the pump does not have any water sensitive equipment. Water will be spilled during any repairs or if a leak occurs.
Use the Clark Pump mounting hole template, see page 34, to mark and drill for the fasteners. 1/4" bolts and Nylock nuts are preferred, but lag bolts are acceptable if they penetrate at least 1" into solid material. Drill pilot holes for lags bolts.
Caution: Tighten the fasteners until there is light pressure on the plastic inserts. The plastic inserts will collapse if over tightened. The 1/4” fender washer must be on top to prevent the mounting brackets from slipping out of the rubber mounts under adverse conditions.
Do not mount the Clark Pump in an area that is subject to sustained temperatures over 120 degrees F.
Component Placement, cont.
Membrane and Housing Assembly
Mount the membrane and housing assembly as close to the Clark Pump and in as cool a
page 25
place as possible. Use the black plastic brackets and heavy duty black tie wraps provided. The plastic tie wraps allow the membrane housing to expand and contract with pressure changes without damage. Mount the membrane preferably at or below the level of the valve body on the Clark Pump to help purge air. It is OK to mount the membrane above the Clark Pump, but keep it as close as possible. The membrane housing is marked for water flow direction and can be mounted at an angle or vertical, but the brine outlet end should be up to help purge air. Route the high pressure tubing to and from the membrane housing as smoothly as possible. Avoid clamping the tubes tightly to any surfaces for sound dampening. "Hanging" the tubes away from a surface with wire ties so that they are neither tight nor completely loose is best.
The product water tube can be connected to either end of the membrane. Plug the unused product port and
Do not mount the membrane in an area that is subject to sustained temperatures over 120 degrees F.
Remote Control Panel
The remote control panel can be mounted anywhere that’s dry and convenient. Use the
page 26
End block B
High pressure port for remote pressure relief valve
Valve block
End block A
Brine out
Stainless steel tube
Composite cylinder and base
Pressure relief valve
Center block
Feed in
Cylinder ring
End Cap
Clark Pump Front View
End block A
High pressure port for remote pressure relief valve
Valve block
End block B
Reset button
Alternate brine out |
Stainless steel tube
Cylinder ring
End cap
Composite cylinder and base
High pressure in
Test port
Center block |
High pressure out |
Clark Pump Back View
page 27
Membrane and Housing |
3 1/2" |
|
|
(8.9cm) |
|
43 1/2" |
|
|
(112.5cm) |
|
|
Clark Pump |
|
|
SPECTRA |
1/4"9 |
23.5cm |
27 1/4" |
5 1/4" |
|
(69.3cm) |
(13.3c) |
|
|
|
page 28
23 1/2"
(59.7cm)
6.5" (16.4cm) with fittings
Feed Pressure |
|
Pressure Relief Valve |
Product Flow |
Open 1/2 to 1 turn only |
|
SPECTRA
GULFSTREAM 400
S a u s a l i t o , C a l i f o r n i a P h o n e ( 4 1 5 ) 3 3 2 - 3 7 8 0 F a x ( 4 1 5 ) 3 3 2 - 8 5 2 7 s p e c t r a w a t e r m a k e r s . c o m
|
Pump 1 |
S a l i n i t y |
B a c k f l u s h |
|
|
|
|||
|
Pump 2 |
Reject |
Good |
|
Basic Operation: |
|
|
|
|
Priming |
|
|
|
|
Open pressure relief valve, run each pump individulally to prime system. |
Pumps will not prime simultaneously. |
Purging
If system has been stored, open pressure relief valve and run for 20 minutes. Attach the product service tube and place into a container. Set product service valve to "service" position, close pressure relief valve and run for 20 minutes. Discard product water.
Startup
Set all valves to "run" position and close pressure relief valve. Start one pump for efficiency, two pumps
for higher production. Check for the "Good" salinity indicator to light. Monitor feed pressure for
Normal Operation
For best performance; run system to fill tanks then turn off pumps. Have the domestic water system perssurized
then start backflush cycle.
Storage
Attach service hoses to unit and lead to a clean container. Open pressure relief valve and start backflush cycle. After backflush cycle, set brine discharge valve to "service" position and then
Allow 1 gallon of flush water into the container and then set the discharge valve to "run". After backflush cycle, mix one container of
and run for 10 minutes, recirculating the solution. Set the brine discharge valve to "run" to empty container. Turn off pumps, return all service valves to "run" position. Service strainer and
14" |
(35.7cm) |
I n t a k e S e r v i c e V a l v e |
P r o d u c t S e r v i c e V a l v e |
B r i n e D i s c h a r g e S e r v i c e V a l v e |
Run |
|
Run |
Off |
Run |
Off |
|
S e r v i c e |
|
I n t a k e S e r v i c e |
|
Brine Discharge Service |
H o s e C o n n e c t i o n |
|
H o s e C o n n e c t i o n |
|
Product Service Tube Connection |
|
Spectra Gulfstream Components Dimensions
15 1/2" |
(39.5cm) |
page 29
High pressure |
Membrane and housing |
Brine outlet |
|
inlet |
Product outlet |
||
|
|||
|
|
1/2" H.P. tube to top H.P. inlet |
|
1/2" H.P. tube from bottom H.P. outlet |
|
|
1/4" Product tube
|
|
|
|
|
|
1/4" Product |
1/4" high pressure tubing from pressure relief valve |
|
|
|
|
|
tube |
|
|
|
|
|
Feed water hose |
Vent |
Pressurized fresh water from |
|
|
|
|
|
|
domestic water system |
|
|
|
|
|
|
|
|
Pump 1 |
Salinity |
Backflush |
|
|
|
|
|
|
|||
|
|
Pump 2 |
R e j e c t |
G o o d |
|
|
|
|
Basic Operation: |
|
|
|
Fresh water tank |
|
|
Priming |
|
|
|
|
|
|
Open pressure relief valve, run each pump individulally to prime system. Pumps will not prime simultaneously. |
|
|||
|
|
Purging |
|
|
|
|
|
|
If system has been stored, open pressure relief valve and run for 20 minutes. Attach the product service |
|
|||
|
|
tube and place into a container. Set product service valve to "service" position, close pressure relief valve |
|
|||
Feed Pressure |
|
and run for 20 minutes. Discard product water. |
|
|
|
|
|
|
Startup |
|
|
|
|
|
|
Set all valves to "run" position and close pressure relief valve. Start one pump for efficiency, two pumps |
|
|||
|
|
for higher production. Check for the "Good" salinity indicator to light. Monitor feed pressure for |
|
|||
|
|
Service |
|
|||
|
|
Normal Operation |
|
|
|
|
|
|
For best performance; run system to fill tanks then turn off pumps. Have the domestic water system perssurized |
|
|||
|
|
then start backflush cycle. |
|
|
||
Pressure Relief Valve |
Product Flow |
Storage |
|
|
|
|
Open1/2to1turnonly |
|
Attach service hoses to unit and lead to a clean container. |
Open pressure relief valve and start backflush cycle. |
|
||
SPECTRA |
|
After backflush cycle, set brine discharge valve to "service" position and then |
|
|||
|
mix one container of |
|
||||
|
|
Allow 1 gallon of flush water into the container and then set the discharge valve to "run". After backflush cycle, |
|
|||
|
GULFST REAM 400 |
and run for 10 minutes, recirculating the solution. |
Set the brine discharge valve to "run" to empty container. |
|
Turn off pumps, return all service valves to "run" position. Service strainer and
|
Purge system on next start up. |
|
|
|
|
|
|
|
Brine discharge |
|
Intake Service Valve |
Product Service Valve |
Brine Discharge Service Valve |
hose |
|
Run |
|
Run |
|
|
Off |
Run |
Off |
|
|
|
Service |
|
|
Sea water intake hose |
|
|
|
|
|
Intake Service |
|
Brine Discharge Service |
|
|
Hose Connection |
|
Hose Connection |
|
|
|
Product ServiceTube Connection |
|
|
|
Charcoal |
|
|
|
|
filter |
|
|
|
|
|
|
Service hoses |
|
Intake |
Strainer |
|
|
|
|
|
|
|
|
|
|
|
Cleaning |
|
|
|
|
container |
|
Gulfstream 400 Plumbing Diagram
Product water from membrane |
Product water to tank |
Brine discharge to |
Brine discharge |
from Clark Pump |
Right Side |
Sea water inlet from straine |
Left Side |
Gulfstream 400 Plumbing Connections
page 30
Bolt mounting |
Lag or screw |
|
mounting |
||
|
||
|
1/4" SS Bolt |
|
|
Caution: Do not over tighten, |
|
|
the plastic inserts may crush. |
# 14 sheet metal screw
1/4" SS Fender Washer
Plastic Insert
Center Block |
Plastic washers |
1/4" SS Fender Washer
Plastic Insert
Rubber Grommet
|
Rubber |
Grommet |
|
Mount Bolts |
|
|
|
Mounting Surface |
|
Plastic washers |
Bottom view
Aluminum brackets
Note: Remove aluminum mounts before shipping the Clark Pump.
Clark Pump Mounting
page 31
Spectra High Pressure Tube Fitting Assembly
Carefully fit and measure the tubing before cutting with a sharp razor knife or hose cutter and remove any burrs. Minimum tubing bend radius is 6”. Route tubing away from excessive heat sources and secure from vibration and chafe. Have at least one shallow bend in a tube assembly after it is installed.
Refer to figure 1. If a fitting has been dissembled, reassemble as illustrated. The notch on the ferrule must engage the inside of the nut properly for the nut to seat down fully. Once the tube is inserted the ferrule and nut will naturally align.
Refer to figure 2. Insert tube fully into the fitting, it should go in 0.9”. Tighten the nut finger tight while moving the tube around to prevent binding. One thread should be showing under the nut. Secure the tube so it won’t back out when tightening.
Refer to figure 3. Use 13/16” wrench to hold a straight body fitting or a 3/4” wrench for a 90º body, and a 7/8” wrench for the nut. Hold the body, recheck the tube insertion, then tighten the nut
The tube connectors can be disconnected and
Index mark |
|
|
Straight thread |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Figure 1.
Nut |
Ferrule |
Body |
Cut tube square
Figure 2. |
Black high pressure tubing |
|
Figure 3.
Tighten
No threads showing |
Insert tube 0.9" until it stops
page 32
|
Spacer |
|
|
Grab Ring |
|
Body |
Nut |
|
|
|
|
|
|
Tubing |
Step 1: |
Dissemble fitting components |
|
|
|
1/2" max |
Step 2:
Step 3: Gently fit the tube into the body and loosely thread on the nut.
Step 4: Push the tube into the body until it bottoms out then hand tighten the nut.
DO NOT OVER TIGHTEN!
1/4" Tube Fitting Assembly
page 33
Drill 1/4" holes to start cut out
Top
Spectra Remote Control Panel
Cut Out Template
Cut line
4.750"
Drill for # 4 screws
to mount panelDrill for 1/4" bolts, 1/4" lags,
# 14 SM Screws
Back
1.750"
Under |
4.000" |
Clark Pump |
Under |
Cylinder |
Mounting Template |
Cylinder |
|
|
|
||
|
|
Caution! Check templates for accuracy before drilling |
|
|
|
6.500" |
|
|
|
Front |
|
page 34
ELECTRICAL
See the wiring diagram on the following page.
A robust electrical connection is crucial to the proper operation of the system. Examine your electrical system carefully to determine the proper place from which to draw power. If power will be drawn from a breaker on the main distribution panel, make sure the panel has at least #8 gauge wire for panel positive feed and ground. Check all connections for tightness, including the ground buss. Refer to the wiring diagrams on this page and
page 36.
If the panel is inadequate or too far away, a direct connection to a battery is possible. An
Power module wiring: For 12VDC use a 25 amp breaker and from the power source to the module and #10 gauge to 10 feet, #8 gauge to 18 feet, and #6 gauge to 28 feet. For 24 VDC systems use a 15 amp breaker and #12 gauge to 10 feet, #10 gauge to 18 feet, and #8 gauge to 28 feet.
Use marine grade insulated
Follow the pattern of the original factory wiring when adding a new circuit. When adding a new wire to an existing wire bundle, remove all the old tie wraps where possible, run the new wire, then tie at even intervals. This will result in a clean and "original" look.
Connect the power wiring to the studs on the face of the power control module, positive to positive as marked.
Troubleshooting Guide
|
Power feed wire size |
Main Electrical Panel |
|
|
|
|
|
|
for 12VDC systems: |
|
|
Spectra Power Control Module |
25 amp breaker required |
|
|
|
# 10 gauge to 10 feet |
|
|
|
# 8 gauge to 18 feet |
|
|
|
# 6 gauge to 28 feet |
|
|
|
|
15 or 25 amp breaker |
|
+ DC Input |
|
Positive buss |
|
|
|
|
|
- DC Ground |
|
|
|
Pump 1 |
|
|
|
Pump 2 |
Power feed wire size |
|
|
for 24VDC systems: |
|
|
|
|
|
|
|
|
15 amp breaker required |
|
|
Diversion Valve |
# 12 gauge to 10 feet |
Negtive buss |
|
|
|
||
|
# 10 gauge to 18 feet |
To battery, #8 gauge or larger_ |
|
|
# 8 gauge to 28 feet |
+ |
|
|
|
|
Gulfstream 400 Power Control Module Wiring Diagram
page 35
page 36
Power Control Module enclosure bottom |
Underside of enclosure top |
|
|
|
||
|
|
|
|
|||
|
|
|
|
Spectra Control Panel |
|
|
|
|
|
|
|
For 200c models omit |
|
|
|
|
|
|
|
|
|
|
|
RED |
|
|
pump 2 and both fans |
|
|
|
|
|
|
|
|
|
|
GRN |
|
|
|
|
|
|
BLK |
|
|
Connect fan wiring with |
|
|
Adjusting pot |
|
|
|
|
|
|
BLU |
|
|
feed pump connections |
|
|
|
|
GRY |
|
|
|
Salinity control |
+ DC Input |
Backflush timer |
WHT |
Cooling fan |
Cooling fan |
|
25 amp |
|
|
|
|
||
PC board |
|
ORG |
|
|
|
|
|
|
|
|
|
||
|
|
|
YEL |
|
|
|
|
|
|
PUR |
|
|
|
|
|
|
BRN |
|
|
|
RCA jack |
|
- DC Input |
|
|
|
|
|
|
|
|
Pump 1 |
Pump 2 |
|
|
|
+ Pump 1 |
|
|
|
|
|
Pump 1 relay |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
- Pump 1 |
|
|
|
|
|
|
|
|
|
|
Solenoid wire size |
|
|
+ Pump 2 |
|
|
|
for 12 or 24VDC |
|
|
|
|
|
|
systems: #16 gauge |
|
Pump 2 relay |
- Pump 2 |
|
Power feed wire size |
Backflush valve |
Diversion valve |
|
|
for 12VDC systems: |
||||
|
|
|
|
|
|
|
|
|
+ Backflush |
|
25 amp breaker required |
|
|
|
|
|
|
|
|
|
|
|
valve |
|
# 10 gauge to 10 feet |
|
|
|
|
- Backflush |
|
# 8 gauge to 18 feet |
|
|
|
|
|
# 6 gauge to 28 feet |
|
|
|
|
|
valve |
|
|
|
|
|
|
|
|
|
|
|
|
Backflush relay |
+ Diversion |
|
Power feed wire size |
|
|
|
|
valve |
|
|
|
|
|
|
|
|
for 24VDC systems: |
|
|
|
|
- Diversion |
|
15 amp breaker required |
|
|
|
|
valve |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
# 12 gauge to 10 feet |
|
|
|
|
|
|
# 10 gauge to 18 feet |
|
|
|
All wires enter through enclosure bottom |
|
# 8 gauge to 28 feet |
|
|
|
|
+ |
|
|
|
|
|
|
25 amp DC power input |
|
|
|
|
|
|
_ |
|
|
|
|
|
|
|
Do not cut probe wire and run seperately from other wires to avoid interference |
|
|
Spectra Gulfstream 400 Wiring Diagram
For instructions on repairing the Clark Pump, please contact Spectra Watermakers.
SYMPTOM |
CAUSE |
REMEDY |
*Feed pump start but shuts down on |
*Clark Pump stalled. |
*Push in reset button |
high pressure. (over 100 psi) |
*Brine service valve in mid (off) |
*Open brine service valve |
|
position |
|
|
*Service |
|
*System does not run |
*Tripped breaker or blown fuse. |
*Check wiring to motor. |
|
*Pump power relay faulty. |
*Replace relay. |
|
*Motor burned out. |
*Replace motor. |
*Feed pump runs but no fresh |
*Check system, then open |
|
water flow. |
*Pressure relief valves open. |
*Close relief valves. |
|
*Pump |
*Open relief valve to purge air, |
|
|
then close. |
*Lower water production, high amp |
*Service |
|
draw, high feed pressure. |
|
|
*Low water production, normal to |
*Relief valve partially open. |
*Tighten hand tight only. |
low amp draw, normal to low feed |
*Leak in high pressure hoses or |
*Tighten or replace fitting and/or |
pressure. |
fittings. |
hose. |
|
*Internal leakage in the Clark |
*See Clark Pump Repair Manual. |
|
Pump. |
|
|
*Debris in feed pump check valves. |
*Disassemble pump head and |
|
|
clean. |
*Water production normal, feed |
*Cold sea water temperature. |
*Normal operation. |
pressure high, high amp draw. |
*Membrane starting to foul. |
*Clean membrane. |
*Water production normal, feed |
*Warm sea water temperature. |
*Normal operation. |
pressure low, amp draw low. |
|
|
*Water production high, poor |
*Seal failure inside membrane |
*Disassemble and check for bad |
water quality. |
housing. |
seals and/or corrosion. |
|
*Membrane failure. |
*Replace membrane. |
*Asymmetrical pressure and flow |
*Reversing valve seal leaking. |
*See Clark Pump Repair Manual. |
readings between shifts. |
*Shaft seal leaking. |
|
|
*Pin seal leaking. |
|
|
*Check valve leaking. |
|
|
*Scored cylinder wall. |
|
|
*Piston seal leaking. |
|
page 37
LIMITED WARRANTY
Spectra Watermakers warrants to the original purchaser that all components, except for the Clark Pump, are free from any defects in material and/or workmanship for a period of one year from the date of purchase.
Spectra Watermakers warrants to the original purchaser that the Clark Pump is free from any defects in material and/or workmanship with a limited life time warranty from the date of purchase.
If any such defect is discovered and reported within the warranty period, Spectra Watermakers will repair or replace the affected component free of charge, subject to verification of the defect or malfunction upon inspection at Spectra Watermakers. The customer is responsible for all costs incurred for removal, shipping to and from our facility, and reinstallation.
This warranty does not apply to malfunctions or physical damage resulting from abuse, neglect, accident, alteration, modification, feed pump substitution, submersion, shipping damage, hoses and fittings not supplied by Spectra Watermakers, improper membrane maintenance, oil and chlorine damage, use of sodium bisulfite or sodium metabisulfite, or improper installation. Consumables such as
Spectra Watermakers, under no circumstances, shall be liable for any consequential damages arising out of, or in any way connected with, the failure of the product to perform as set forth herein. This limited warranty is in lieu of any and all other expressed or implied warranties.
Spectra Watermakers reserves the right to make changes or improvements on its products without incurring the obligation to retrofit such changes or improvements on previously manufactured units.
If a possible warranty problem is detected, please contact one of our service representatives before any disassembly for assistance, and to receive a Return Authorization number if a component is to be returned for a warranty repair.
Proof of purchase, return address, and a Return Authorization number are required for all warranty repairs.
Ship to:
Spectra Watermakers
298 Harbor Drive |
Sausalito, CA 94965 |
Phone 415.332.3780 |
Fax 415.332.8527 |
page 38